OSLD energy response performance and dose accuracy at 24-1250 keV

Comparison with TLD-100H and TLD-100

A. B A Kadir, W. Priharti, Supian Samat, M. T. Dolah

Research output: Chapter in Book/Report/Conference proceedingConference contribution

4 Citations (Scopus)

Abstract

OSLD was evaluated in terms of energy response and accuracy of the measured dose in comparison with TLD-100H and TLD-100. The OSLD showed a better energy response performance for Hp(10) whereas for Hp(0.07), TLD-100H is superior than the others. The OSLD dose accuracy is comparable with the other two dosimeters since it fulfilled the requirement of the ICRP trumpet graph analysis.

Original languageEnglish
Title of host publicationAIP Conference Proceedings
Pages108-114
Number of pages7
Volume1571
DOIs
Publication statusPublished - 2013
Event2013 UKM Faculty of Science and Technology Post-Graduate Colloquium - Selangor
Duration: 3 Jul 20134 Jul 2013

Other

Other2013 UKM Faculty of Science and Technology Post-Graduate Colloquium
CitySelangor
Period3/7/134/7/13

Fingerprint

dosage
dosimeters
requirements
energy

Keywords

  • Energy response
  • H(0.07)
  • H(10)
  • ICRP trumpet graph
  • OSL
  • TLD-100
  • TLD-100H

ASJC Scopus subject areas

  • Physics and Astronomy(all)

Cite this

Kadir, A. B. A., Priharti, W., Samat, S., & Dolah, M. T. (2013). OSLD energy response performance and dose accuracy at 24-1250 keV: Comparison with TLD-100H and TLD-100. In AIP Conference Proceedings (Vol. 1571, pp. 108-114) https://doi.org/10.1063/1.4858638

OSLD energy response performance and dose accuracy at 24-1250 keV : Comparison with TLD-100H and TLD-100. / Kadir, A. B A; Priharti, W.; Samat, Supian; Dolah, M. T.

AIP Conference Proceedings. Vol. 1571 2013. p. 108-114.

Research output: Chapter in Book/Report/Conference proceedingConference contribution

Kadir, ABA, Priharti, W, Samat, S & Dolah, MT 2013, OSLD energy response performance and dose accuracy at 24-1250 keV: Comparison with TLD-100H and TLD-100. in AIP Conference Proceedings. vol. 1571, pp. 108-114, 2013 UKM Faculty of Science and Technology Post-Graduate Colloquium, Selangor, 3/7/13. https://doi.org/10.1063/1.4858638
Kadir, A. B A ; Priharti, W. ; Samat, Supian ; Dolah, M. T. / OSLD energy response performance and dose accuracy at 24-1250 keV : Comparison with TLD-100H and TLD-100. AIP Conference Proceedings. Vol. 1571 2013. pp. 108-114
@inproceedings{3947375b808645bdb70721edee6f5725,
title = "OSLD energy response performance and dose accuracy at 24-1250 keV: Comparison with TLD-100H and TLD-100",
abstract = "OSLD was evaluated in terms of energy response and accuracy of the measured dose in comparison with TLD-100H and TLD-100. The OSLD showed a better energy response performance for Hp(10) whereas for Hp(0.07), TLD-100H is superior than the others. The OSLD dose accuracy is comparable with the other two dosimeters since it fulfilled the requirement of the ICRP trumpet graph analysis.",
keywords = "Energy response, H(0.07), H(10), ICRP trumpet graph, OSL, TLD-100, TLD-100H",
author = "Kadir, {A. B A} and W. Priharti and Supian Samat and Dolah, {M. T.}",
year = "2013",
doi = "10.1063/1.4858638",
language = "English",
isbn = "9780735411999",
volume = "1571",
pages = "108--114",
booktitle = "AIP Conference Proceedings",

}

TY - GEN

T1 - OSLD energy response performance and dose accuracy at 24-1250 keV

T2 - Comparison with TLD-100H and TLD-100

AU - Kadir, A. B A

AU - Priharti, W.

AU - Samat, Supian

AU - Dolah, M. T.

PY - 2013

Y1 - 2013

N2 - OSLD was evaluated in terms of energy response and accuracy of the measured dose in comparison with TLD-100H and TLD-100. The OSLD showed a better energy response performance for Hp(10) whereas for Hp(0.07), TLD-100H is superior than the others. The OSLD dose accuracy is comparable with the other two dosimeters since it fulfilled the requirement of the ICRP trumpet graph analysis.

AB - OSLD was evaluated in terms of energy response and accuracy of the measured dose in comparison with TLD-100H and TLD-100. The OSLD showed a better energy response performance for Hp(10) whereas for Hp(0.07), TLD-100H is superior than the others. The OSLD dose accuracy is comparable with the other two dosimeters since it fulfilled the requirement of the ICRP trumpet graph analysis.

KW - Energy response

KW - H(0.07)

KW - H(10)

KW - ICRP trumpet graph

KW - OSL

KW - TLD-100

KW - TLD-100H

UR - http://www.scopus.com/inward/record.url?scp=84897783548&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84897783548&partnerID=8YFLogxK

U2 - 10.1063/1.4858638

DO - 10.1063/1.4858638

M3 - Conference contribution

SN - 9780735411999

VL - 1571

SP - 108

EP - 114

BT - AIP Conference Proceedings

ER -