Kaedah relatif untuk penentuan nilai pembetulan medan bebas mikrofon pengukuran bunyi

Research output: Contribution to journalArticle

Abstract

A free-field correction value is a value used to convert the sound pressure response of measuring microphones during calibration from pressure field response to free-field response. A relative comparison method for determination of this correction value has been developed as an alternative to the established absolute methods. Pressure response has been compared relatively with free-field response. Microphone's response was obtained by comparison with laboratory standard microphone where the free-field correction value is known. Both microphones were sequentially exposed to the same sound pressure level in free field and pressure field and AC voltages were measured. The free field is generated inside an anechoic chamber and pressure field using a sound calibrator. The correction values of two measuring microphones with known free-field correction value given by the manufacturer were measured for method validation. Both results were compared and show the agreement within the uncertainties and this proved the relative method can be applied.

Original languageUndefined/Unknown
Pages (from-to)25-32
Number of pages8
JournalJurnal Teknologi (Sciences and Engineering)
Volume64
Issue number1
DOIs
Publication statusPublished - Sep 2013

Keywords

  • Free-field calibration of microphone
  • Free-field correction
  • Measuring microphone
  • Pressure calibration of microphone
  • Relative method

ASJC Scopus subject areas

  • Engineering(all)

Cite this

Kaedah relatif untuk penentuan nilai pembetulan medan bebas mikrofon pengukuran bunyi. / Wan Aziz, Wan Salleh; Jamaludin, Nordin; Mohd Nor, Mohd. Jailani.

In: Jurnal Teknologi (Sciences and Engineering), Vol. 64, No. 1, 09.2013, p. 25-32.

Research output: Contribution to journalArticle

@article{1e3c76e226ec4bc0b837b3cfb30b0f87,
title = "Kaedah relatif untuk penentuan nilai pembetulan medan bebas mikrofon pengukuran bunyi",
abstract = "A free-field correction value is a value used to convert the sound pressure response of measuring microphones during calibration from pressure field response to free-field response. A relative comparison method for determination of this correction value has been developed as an alternative to the established absolute methods. Pressure response has been compared relatively with free-field response. Microphone's response was obtained by comparison with laboratory standard microphone where the free-field correction value is known. Both microphones were sequentially exposed to the same sound pressure level in free field and pressure field and AC voltages were measured. The free field is generated inside an anechoic chamber and pressure field using a sound calibrator. The correction values of two measuring microphones with known free-field correction value given by the manufacturer were measured for method validation. Both results were compared and show the agreement within the uncertainties and this proved the relative method can be applied.",
keywords = "Free-field calibration of microphone, Free-field correction, Measuring microphone, Pressure calibration of microphone, Relative method",
author = "{Wan Aziz}, {Wan Salleh} and Nordin Jamaludin and {Mohd Nor}, {Mohd. Jailani}",
year = "2013",
month = "9",
doi = "10.11113/jt.v64.1163",
language = "Undefined/Unknown",
volume = "64",
pages = "25--32",
journal = "Jurnal Teknologi",
issn = "0127-9696",
publisher = "Penerbit Universiti Teknologi Malaysia",
number = "1",

}

TY - JOUR

T1 - Kaedah relatif untuk penentuan nilai pembetulan medan bebas mikrofon pengukuran bunyi

AU - Wan Aziz, Wan Salleh

AU - Jamaludin, Nordin

AU - Mohd Nor, Mohd. Jailani

PY - 2013/9

Y1 - 2013/9

N2 - A free-field correction value is a value used to convert the sound pressure response of measuring microphones during calibration from pressure field response to free-field response. A relative comparison method for determination of this correction value has been developed as an alternative to the established absolute methods. Pressure response has been compared relatively with free-field response. Microphone's response was obtained by comparison with laboratory standard microphone where the free-field correction value is known. Both microphones were sequentially exposed to the same sound pressure level in free field and pressure field and AC voltages were measured. The free field is generated inside an anechoic chamber and pressure field using a sound calibrator. The correction values of two measuring microphones with known free-field correction value given by the manufacturer were measured for method validation. Both results were compared and show the agreement within the uncertainties and this proved the relative method can be applied.

AB - A free-field correction value is a value used to convert the sound pressure response of measuring microphones during calibration from pressure field response to free-field response. A relative comparison method for determination of this correction value has been developed as an alternative to the established absolute methods. Pressure response has been compared relatively with free-field response. Microphone's response was obtained by comparison with laboratory standard microphone where the free-field correction value is known. Both microphones were sequentially exposed to the same sound pressure level in free field and pressure field and AC voltages were measured. The free field is generated inside an anechoic chamber and pressure field using a sound calibrator. The correction values of two measuring microphones with known free-field correction value given by the manufacturer were measured for method validation. Both results were compared and show the agreement within the uncertainties and this proved the relative method can be applied.

KW - Free-field calibration of microphone

KW - Free-field correction

KW - Measuring microphone

KW - Pressure calibration of microphone

KW - Relative method

UR - http://www.scopus.com/inward/record.url?scp=84883758316&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84883758316&partnerID=8YFLogxK

U2 - 10.11113/jt.v64.1163

DO - 10.11113/jt.v64.1163

M3 - Article

VL - 64

SP - 25

EP - 32

JO - Jurnal Teknologi

JF - Jurnal Teknologi

SN - 0127-9696

IS - 1

ER -