Development of a two channel dynamic phantom for fetal heart rate and pulse oximetry application

M. A M Yahya, Gan Kok Beng

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

A Continuous and real time health observing has become an important subject of our life. A simple and feasible Portable devices has done a major role, especially in the biomedical field. We have implemented and developed with an experimental study for two pulsation tubes simultaneously as a fetal and maternal abdominal vessel based on optical continuous detection. A wide range of frequencies from near-infrared to red light wavelength were generated through a phantom. The reflected signals were acquired from one silicon diode and separated by a demodulator through a USB connection to PC by using an NI DAQ 6351 and analysed using LABVIEW software. The settled simulated vascular system contains from a two independent adjustable tube system, demonstrating the fetal and maternal vessel system in different frequencies simultaneously. The two arterial hearts beat pressure with different frequencies is generated utilizing a two different pre-pressure controllable piston pumps with two handmade automatic adjustable relief valve and two flexible tubes to create a complete artificial model of the fetal and maternal vessel system of a dynamic adjustable beat per minute (B.P.M.). The major achieved results are the reflected RED and IR PPG signals from both top and bottom tube simultaneously from one silicon diode. The system that has designed is convenient with any detection systems for a transabdominal optical PPG system tests before produced to the clinics.

Original languageEnglish
Pages (from-to)1277-1280
Number of pages4
JournalAdvanced Science Letters
Volume23
Issue number2
DOIs
Publication statusPublished - 1 Feb 2017

Fingerprint

Fetal Heart Rate
Oximetry
Heart Rate
Phantom
Diodes
vessel
Mothers
Silicon
Pressure relief valves
Reciprocating pumps
silicon
Demodulators
Tube
Optical systems
Vessel
Pressure
Optical Devices
Health
Beat
Infrared radiation

Keywords

  • Dynamic optical phantom mimicking
  • Fetal heart rate (FHR)
  • Photoplesthysmography (PPG)
  • Reflectance pulse oximetry
  • Transabdominal

ASJC Scopus subject areas

  • Health(social science)
  • Computer Science(all)
  • Education
  • Mathematics(all)
  • Environmental Science(all)
  • Engineering(all)
  • Energy(all)

Cite this

Development of a two channel dynamic phantom for fetal heart rate and pulse oximetry application. / Yahya, M. A M; Kok Beng, Gan.

In: Advanced Science Letters, Vol. 23, No. 2, 01.02.2017, p. 1277-1280.

Research output: Contribution to journalArticle

@article{b0a85118929a48209509b89cb92d02db,
title = "Development of a two channel dynamic phantom for fetal heart rate and pulse oximetry application",
abstract = "A Continuous and real time health observing has become an important subject of our life. A simple and feasible Portable devices has done a major role, especially in the biomedical field. We have implemented and developed with an experimental study for two pulsation tubes simultaneously as a fetal and maternal abdominal vessel based on optical continuous detection. A wide range of frequencies from near-infrared to red light wavelength were generated through a phantom. The reflected signals were acquired from one silicon diode and separated by a demodulator through a USB connection to PC by using an NI DAQ 6351 and analysed using LABVIEW software. The settled simulated vascular system contains from a two independent adjustable tube system, demonstrating the fetal and maternal vessel system in different frequencies simultaneously. The two arterial hearts beat pressure with different frequencies is generated utilizing a two different pre-pressure controllable piston pumps with two handmade automatic adjustable relief valve and two flexible tubes to create a complete artificial model of the fetal and maternal vessel system of a dynamic adjustable beat per minute (B.P.M.). The major achieved results are the reflected RED and IR PPG signals from both top and bottom tube simultaneously from one silicon diode. The system that has designed is convenient with any detection systems for a transabdominal optical PPG system tests before produced to the clinics.",
keywords = "Dynamic optical phantom mimicking, Fetal heart rate (FHR), Photoplesthysmography (PPG), Reflectance pulse oximetry, Transabdominal",
author = "Yahya, {M. A M} and {Kok Beng}, Gan",
year = "2017",
month = "2",
day = "1",
doi = "10.1166/asl.2017.8390",
language = "English",
volume = "23",
pages = "1277--1280",
journal = "Advanced Science Letters",
issn = "1936-6612",
publisher = "American Scientific Publishers",
number = "2",

}

TY - JOUR

T1 - Development of a two channel dynamic phantom for fetal heart rate and pulse oximetry application

AU - Yahya, M. A M

AU - Kok Beng, Gan

PY - 2017/2/1

Y1 - 2017/2/1

N2 - A Continuous and real time health observing has become an important subject of our life. A simple and feasible Portable devices has done a major role, especially in the biomedical field. We have implemented and developed with an experimental study for two pulsation tubes simultaneously as a fetal and maternal abdominal vessel based on optical continuous detection. A wide range of frequencies from near-infrared to red light wavelength were generated through a phantom. The reflected signals were acquired from one silicon diode and separated by a demodulator through a USB connection to PC by using an NI DAQ 6351 and analysed using LABVIEW software. The settled simulated vascular system contains from a two independent adjustable tube system, demonstrating the fetal and maternal vessel system in different frequencies simultaneously. The two arterial hearts beat pressure with different frequencies is generated utilizing a two different pre-pressure controllable piston pumps with two handmade automatic adjustable relief valve and two flexible tubes to create a complete artificial model of the fetal and maternal vessel system of a dynamic adjustable beat per minute (B.P.M.). The major achieved results are the reflected RED and IR PPG signals from both top and bottom tube simultaneously from one silicon diode. The system that has designed is convenient with any detection systems for a transabdominal optical PPG system tests before produced to the clinics.

AB - A Continuous and real time health observing has become an important subject of our life. A simple and feasible Portable devices has done a major role, especially in the biomedical field. We have implemented and developed with an experimental study for two pulsation tubes simultaneously as a fetal and maternal abdominal vessel based on optical continuous detection. A wide range of frequencies from near-infrared to red light wavelength were generated through a phantom. The reflected signals were acquired from one silicon diode and separated by a demodulator through a USB connection to PC by using an NI DAQ 6351 and analysed using LABVIEW software. The settled simulated vascular system contains from a two independent adjustable tube system, demonstrating the fetal and maternal vessel system in different frequencies simultaneously. The two arterial hearts beat pressure with different frequencies is generated utilizing a two different pre-pressure controllable piston pumps with two handmade automatic adjustable relief valve and two flexible tubes to create a complete artificial model of the fetal and maternal vessel system of a dynamic adjustable beat per minute (B.P.M.). The major achieved results are the reflected RED and IR PPG signals from both top and bottom tube simultaneously from one silicon diode. The system that has designed is convenient with any detection systems for a transabdominal optical PPG system tests before produced to the clinics.

KW - Dynamic optical phantom mimicking

KW - Fetal heart rate (FHR)

KW - Photoplesthysmography (PPG)

KW - Reflectance pulse oximetry

KW - Transabdominal

UR - http://www.scopus.com/inward/record.url?scp=85015748392&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85015748392&partnerID=8YFLogxK

U2 - 10.1166/asl.2017.8390

DO - 10.1166/asl.2017.8390

M3 - Article

VL - 23

SP - 1277

EP - 1280

JO - Advanced Science Letters

JF - Advanced Science Letters

SN - 1936-6612