Determination of Fano factor and pre-amplifier noise from the measurement of energy resolution of a HPGe detector

Supian Samat, W. Priharti

Research output: Contribution to journalArticle

4 Citations (Scopus)

Abstract

A HPGe detector and γ-ray multinuclide standard solution were used to get several data pairs of photon energy E and detector's energy resolution FWHM. A series of graph of FWHM<sup>2</sup> versus E was plotted to determine the detector's Fano factor and pre-amplifier noise, which yielded the mean values of 0.123±0.003 and (7.448±0.011) × 10<sup>-1</sup> keV, respectively. The Fano factor value agrees within 95% confident interval with the mostly quoted value of 0.13 for semiconductor detectors. The obtained preamplifier noise is in agreement with typical manufacturer's specification.

Original languageEnglish
Pages (from-to)761-764
Number of pages4
JournalSains Malaysiana
Volume44
Issue number5
Publication statusPublished - 1 May 2015

Fingerprint

amplifiers
detectors
energy
preamplifiers
specifications
rays
intervals
photons

Keywords

  • Fano factor
  • FWHM
  • Gamma-ray spectrometry
  • Pre-amplifier noise

ASJC Scopus subject areas

  • General

Cite this

Determination of Fano factor and pre-amplifier noise from the measurement of energy resolution of a HPGe detector. / Samat, Supian; Priharti, W.

In: Sains Malaysiana, Vol. 44, No. 5, 01.05.2015, p. 761-764.

Research output: Contribution to journalArticle

@article{2a913a08b73d47628cd2ba26699c2de4,
title = "Determination of Fano factor and pre-amplifier noise from the measurement of energy resolution of a HPGe detector",
abstract = "A HPGe detector and γ-ray multinuclide standard solution were used to get several data pairs of photon energy E and detector's energy resolution FWHM. A series of graph of FWHM2 versus E was plotted to determine the detector's Fano factor and pre-amplifier noise, which yielded the mean values of 0.123±0.003 and (7.448±0.011) × 10-1 keV, respectively. The Fano factor value agrees within 95{\%} confident interval with the mostly quoted value of 0.13 for semiconductor detectors. The obtained preamplifier noise is in agreement with typical manufacturer's specification.",
keywords = "Fano factor, FWHM, Gamma-ray spectrometry, Pre-amplifier noise",
author = "Supian Samat and W. Priharti",
year = "2015",
month = "5",
day = "1",
language = "English",
volume = "44",
pages = "761--764",
journal = "Sains Malaysiana",
issn = "0126-6039",
publisher = "Penerbit Universiti Kebangsaan Malaysia",
number = "5",

}

TY - JOUR

T1 - Determination of Fano factor and pre-amplifier noise from the measurement of energy resolution of a HPGe detector

AU - Samat, Supian

AU - Priharti, W.

PY - 2015/5/1

Y1 - 2015/5/1

N2 - A HPGe detector and γ-ray multinuclide standard solution were used to get several data pairs of photon energy E and detector's energy resolution FWHM. A series of graph of FWHM2 versus E was plotted to determine the detector's Fano factor and pre-amplifier noise, which yielded the mean values of 0.123±0.003 and (7.448±0.011) × 10-1 keV, respectively. The Fano factor value agrees within 95% confident interval with the mostly quoted value of 0.13 for semiconductor detectors. The obtained preamplifier noise is in agreement with typical manufacturer's specification.

AB - A HPGe detector and γ-ray multinuclide standard solution were used to get several data pairs of photon energy E and detector's energy resolution FWHM. A series of graph of FWHM2 versus E was plotted to determine the detector's Fano factor and pre-amplifier noise, which yielded the mean values of 0.123±0.003 and (7.448±0.011) × 10-1 keV, respectively. The Fano factor value agrees within 95% confident interval with the mostly quoted value of 0.13 for semiconductor detectors. The obtained preamplifier noise is in agreement with typical manufacturer's specification.

KW - Fano factor

KW - FWHM

KW - Gamma-ray spectrometry

KW - Pre-amplifier noise